    Work of breathing in adaptive pressure control continuous mandatory ventilation.

    • Eduardo Mireles-Cabodevila and Robert L Chatburn.
    • Respiratory Institute, Cleveland Clinic, Cleveland, OH, USA. mirelee@uams.edu
    • Resp Care. 2009 Nov 1;54(11):1467-72.

    BackgroundAdaptive pressure control is a mode of mechanical ventilation where inflation pressure is adjusted by the ventilator to achieve a target tidal volume (VT). This means that as patient effort increases, inflation pressure is reduced, which may or may not be clinically appropriate. The purpose of this study was to evaluate the relationship between ventilator work output and patient effort in adaptive pressure control.MethodsA lung simulator (ASL 5000) was set at compliance=0.025 L/cm H2O and resistance=10 cm H2O/L/s. Muscle pressure (Pmus) was a sine wave (20% inspiration, 5% hold, 20% release) that increased from 0-25 cm H2O in steps of 5 cm H2O. The adaptive-pressure-control modes tested were: AutoFlow (Dräger Evita XL), VC+ (Puritan Bennett 840), APV (Hamilton Galileo), and PRVC (Siemens Servo-i and Siemens Servo 300). The target VT was set at 320 mL (Pmus=15 cm H2O, inspiratory pressure=0 cm H2O) to allow delivery of a realistic VT as the simulated patient demanded more volume. All measurements were obtained from the simulator.ResultsPatient work of breathing (patient WOB) increased from 0 J/L to 1.88 J/L through the step increase in Pmus. Target VT was maintained as long as Pmus was below 10 cm H2O. VT then increased linearly with increased Pmus. The ventilators showed 3 patterns of behavior in response to an increase in Pmus: (1) ventilator WOB gradually decreased to 0 J/L as Pmus increased; (2) ventilator WOB decreased at the same rate as Pmus increased but plateaued at Pmus=10 cm H2O by delivering a minimum inspiratory pressure level of 6 cm H2O; (3) ventilator WOB decreased as in patterns 1 and 2 to Pmus=10 cm H2O, but then decreased at a much slower rate.ConclusionsAdaptive-pressure-control algorithms differ between ventilators in their response to increasing patient effort. Notably, some ventilators allow the patient to assume all of the WOB, and some provide a minimum level of WOB regardless of patient effort.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.