• Arch. Bronconeumol. · May 2006

    [Transcutaneous measurement of partial pressure of carbon dioxide and oxygen saturation: validation of the SenTec monitor].

    • Ch Domingo, E Canturri, M Luján, A Moreno, H Espuelas, and A Marín.
    • Servei de Pneumologia, Corporació Sanitària Parc Taulí, Sabadell, Barcelona, España. cdomingo@cspt.es
    • Arch. Bronconeumol. 2006 May 1;42(5):246-51.

    ObjectiveTo validate a monitor for transcutaneous measurement of oxygen saturation (SpO2) and partial pressure of carbon dioxide (TcPCO2).Patients And MethodsThis observational study included 140 Caucasian nonsmokers without jaundice. Patients underwent forced spirometry, measurement of SpO2 and TcPCO2 with the SenTec monitor, and arterial blood gas analysis (readings with 2 devices) during the stabilization phase of the monitor. In the statistical analysis, values from the 2 devices for measuring arterial blood gases were compared by mean differences for PaCO2 and oxygen saturation (SaO2). The arithmetic mean of the 2 blood gas measurements was calculated and relations between them and the SpO2 and TcPCO2 were assessed by the Pearson correlation coefficient (r) and the intraclass correlation coefficient (ICC) as a measure of agreement. Bland-Altman analysis was used to test data dispersion.ResultsTen patients were excluded due to a systematic error in the gas calibrator. The mean (SD) time to stabilization of the monitor before reading was 13.9 (2.4) minutes. The forced expiratory volume in the first second was greater than 80% in 40 patients, between 60% and 79% in 23, between 40% and 59% in 30, and less than 40% in 37. The mean (SD) differences between arterial blood gas measurements were 0.28 (1.0) mm Hg for PaCO2, -0.06% (0.86%) for SaO2, and -0.9 (2.7) mm Hg for PaO2. In the tests for correlation and agreement, r was 0.74 and ICC was 0.73 for SaO2 and SpO2; r was 0.92 and ICC was 0.92 for PaCO2 and TcPCO2. The subgroup analyses did not show any noteworthy differences. The Bland Altman analysis showed no significant dispersion. It was observed that the SenTec monitor underestimated oxygen saturation values by around 1% with respect to SaO2 and overestimated carbon dioxide pressure by 1 mm Hg with respect to PaCO2 values.ConclusionsThe stabilization time recommended for the SenTec monitor before taking a reading is 20 minutes. The overestimates and underestimates by the monitor are not clinically relevant. Finally, the values for SpO2 and TcPCO2 measured by the validated monitor are reliable.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.