A clinical evaluation of the Codman MicroSensor for intracranial pressure monitoring.
- D F Signorini, A Shad, I R Piper, and P F Statham.
- Department of Clinical Neurosciences, University of Edinburgh, UK. dfs@skull.dcn.ed.ac.uk
- Br J Neurosurg. 1998 Jun 1;12(3):223-7.
AbstractThe strain-gauge Codman MicroSensor intracranial pressure (ICP) transducer has shown consistently good laboratory performance. To assess the practical performance of the system in patients following acute brain injury, 10 patients were fitted with a MicroSensor and a second ICP monitor. In five cases this was a fibre-optic transducer and in five cases an intraventricular fluid-filled device. Paired ICP values were recorded every 5 min. ICP values ranged from 0 to 31 mmHg. Altman-Bland plots showed that individual readings could differ by as much as 9 mmHg. Further analysis showed that much of this disagreement could be explained by a constant offset on each occasion. Comparison traces of ICP in individual patients show high agreement in timing and size of changes. The unexplained constant offset leads to uncertainty about the true ICP. Treatment decisions are often based upon absolute levels of ICP and patient care may therefore differ depending upon the monitor used.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.