Arch Orthop Trauma Surg · Sep 2010
Comparative StudyReduction of radiation dose during radiofrequency denervation of the lumbar facet joints using the new targeting system SabreSource: a prospective study in 20 patients.
- Dirk Proschek, Konstantinos Kafchitsas, Michael Rauschmann, Andreas Kurth, Thomas Vogl, and Florian Geiger.
- Department of Trauma and Orthopaedic Surgery, University Hospital, University of Mainz, Langenbeckstrasse 1, Mainz, Germany. proschek@arcor.de
- Arch Orthop Trauma Surg. 2010 Sep 1;130(9):1103-10.
Background And AimsFacet joint denervation is a frequently performed technique to treat facet joint syndrome. Most often this technique is used under fluoroscopic guidance implicating high radiation doses for both patients and surgeons. This prospective study was performed to evaluate the effectiveness in reducing radiation dose during radiofrequency ablation therapy of the lumbar facet joints and to evaluate the feasibility and possibilities of the new real time image guidance system SabreSource.Materials And MethodsAs much as 20 consecutive patients with radiofrequency ablation therapy of the facet joints L4 to S1 were included. Ten patients were treated by fluoroscopic control alone; the following 10 patients were treated with the SabreSource image guidance system. A total of 40 thermal ablations to the facet joints were performed. Each patient was given one thermal ablation on both sides of the vertebral segment, either to the facet joints of L4-L5 or of L5-S1. Pain, according to the visual analogue scale (VAS), was documented before and 6 h after the intervention. Radiation dose, time of radiation and the number of shots needed to place the radiofrequency cannula were recorded.ResultsNo complications occurred. Before therapy, the mean VAS in all patients was 7.6 (range 6-10). After therapy the mean VAS in all patients was 3.4 (range 0-5). Compared to the fluoroscopy-guided thermal ablation therapy the SabreSource system significantly reduced the number of fluoroscopy exposures (reduction 23.53%, p = 0.02), the time of radiation exposure (reduction 21.2%, p = 0.03) and the mean entrance surface dose (reduction 30.46%, p = 0.01).ConclusionThe SabreSource System reduces radiation exposure and radiation dose in the radiofrequency denervation therapy of the lumbar facet joints and can be applied for other minimally invasive techniques.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.