• Intensive care medicine · May 1996

    Multicenter Study Comparative Study Clinical Trial

    Clinical evaluation of a multiparameter intra-arterial blood-gas sensor.

    • E Abraham, T J Gallagher, and S Fink.
    • Division of Pulmonary Sciences and Critical Care Medicine, University of Colorado Health Sciences Center, Denver 80262, USA.
    • Intensive Care Med. 1996 May 1;22(5):507-13.

    ObjectiveTo evaluate the in vivo performance of a continuous, intra-arterial, multiparameter blood-gas sensor containing a thermocouple, miniaturized polarographic oxygen electrode, as well as fiberoptic pH and pCO2 sensors.DesignProspective, multicenter study comparing pH, PaCO2, and PaO2 measurements from the intraarterial sensor with those obtained from a conventional, laboratory blood-gas monitor.SettingIntensive care units in three academic medical centers.PatientsAdult ICU patients (n = 26) with a clinical need for a radial artery catheter and frequent monitoring of arterial blood gases for > or = 3 days.InterventionsAll patients had the multiparameter intra-arterial sensor placed through a 20-gauge catheter into the radial artery. Every 4 h, or more frequently if clinically indicated, a blood-gas sample was withdrawn from the radial artery catheter and sent to the laboratory for analysis using a conventional laboratory blood-gas analyzer. Immediately prior to withdrawal of the arterial blood, values for pH, PaCO2, and PaO2 from the multiparameter intra-arterial sensor were recorded.Measurements And Main ResultsThe multiparameter sensor was placed into 26 patients. In 7 of the patients, premature discontinuation of monitoring was necessary because of dampening of the pressure tracing, difficulty in withdrawing blood from the arterial catheter, or bending of the extra-arterial fiberoptic channel owing to the cable being inadequately secured (mean monitoring time in these 7 patients: 40.6 +/- 25.7 h). In the other 19 patients, monitoring was continued until no longer clinically indicated. In these patients, a total of 341 data sets was collected, with the average length of monitoring being 69.9 +/- 37.9 h (range 22.0 to 57.9 h). Comparison of the sensor values with those from the blood-gas analyzer showed bias and precision values of 0.006 and 0.026 for arterial pH - 1.19% and 12.54% for PaO2, and 1.28 mmHg and 2.48 mmHg for PaCO2, respectively. No complications were associated with the intra-arterial sensor.ConclusionsClinical performance of this intra-arterial, multiparameter blood-gas sensor demonstrated stability, consistency, and accuracy comparable to laboratory blood-gas analyzers. The present multiparameter, intravascular blood-gas sensor, when inserted in the radial artery, can provide stable and accurate monitoring of pH, PaCO2, and PaO2 over clinically relevant periods of as long as 6 days in the critical care setting.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…