Ann Fr Anesth Reanim · Jan 1998
[Halogenated anesthetic vaporizers: importance of maintenance regulations].
- C E Klopfenstein, B Walder, A Rohrbach, F X Clergue, and A Forster.
- Division d'anesthésiologie, hôpital universitaire, Genève.
- Ann Fr Anesth Reanim. 1998 Jan 1;17(7):747-9.
AbstractWe report a serious dysfunction of 19 halothane vaporisers Vapor 19.3 (Dräger) which delivered a much higher concentration of agent than indicated on the dial. This inaccuracy was linked to a major corrosion of the inner layers of the vaporiser, with a deposit of zinc bromide and chloride in the bypass channel. The main cause for this dysfunction was the absence of an adequate maintenance of the vaporisers since their purchase 3 years before.