    [Laboratory testing measurement of FIO2 delivered by Boussignac CPAP system with an input of 100% oxygen].

    • F Templier, F Dolveck, M Baer, M Chauvin, and D Fletcher.
    • Samu 92-Smur Garches, hôpital Raymond-Poincaré, 104, boulevard Raymond-Poincaré, 92380, Garches, France. francois.templier@samu.org
    • Ann Fr Anesth Reanim. 2003 Feb 1; 22 (2): 103-7.

    IntroductionCPAP (Continuous Positive Airway Pressure) is one of the treatments of the cardiogenic acute pulmonary edema (cAPE). Among the out-of-hospital used devices, Boussignac CPAP (Vygon) shows a good technical usability, but no analysis of delivered FIO2 is available.ObjectiveTo measure the FIO2 delivered to patient, using Boussignac CPAP with an input of 100 vol % oxygen. Type of study. - Measurement in a laboratory testing.Material And MethodMeasurement of delivered FIO2 in a three parameters environment: expiratory positive airway pressure (EPAP), respiratory rate (RR) and required oxygen flow (ROF). Laboratory test: 100 vol % oxygen input, with a ball-type flow regulator for oxygen 0-30 l x min(-1) (Mediline), Boussignac CPAP valve with manometer (Vygon), Michigan test lung, ventilator César (Taema), oxygen analysis (Servomex) with paramagnetic sensor, with a response time < 10 s, pneumotachographe (Fleisch v 2).ResultsUnder 10 cmH2O EPAP, the required oxygen flow is < or = 30 l x min(-1). Measured FIO2 ranges from 70 to 100 vol %, for a volume per minute < or = 15 l x min(-1) except for a EPAP at 2.5 cmH2O with a rate = 10 c min(-1) and a tidal volume (VT) at 1500 ml where the measured FIO2 is 60 vol %. For a volume per minute > 15 l x min(-1) and < 20 l x min(-1), measured FIO2 ranges from 59 to 83 vol % depending on the variations of RR and VT.ConclusionBoussignac CPAP with 100 vol % oxygen input, delivers high levels of FIO2, especially for volume per minute values usually met in cAPE. The needed oxygen flow is lower than what is usually required by other CPAP flow generators using the venturi effect that may be used in out-of-hospital medical care.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.