• Resp Care · Aug 2005

    Comparative Study

    Albuterol delivery via tracheostomy tube.

    • Christopher M Piccuito and Dean R Hess.
    • Department of Respiratory Care, Massachusetts General Hospital, Boston, Massachusetts 02114, USA.
    • Resp Care. 2005 Aug 1;50(8):1071-6.

    HypothesisAlbuterol delivery through a tracheostomy tube is affected by device (nebulizer vs metered-dose inhaler), interface (mask vs T-piece), bias flow, and humidification.MethodsA lift bar was placed between the chambers of a dual-chambered lung model such that a ventilator triggered simulated spontaneous breathing at a rate of 20 breaths/min, tidal volume of 0.4 L, and inspiratory-expiratory ratio of 1:2. An 8-mm inner diameter cuffed tracheostomy tube was placed through a semi-circular model that simulated a patient's neck. Four conditions of gas flow and humidification were used for the nebulizer experiments: heated aerosol (approximately 30 L/min, approximately 30 degrees C), heated humidity (approximately 30 L/min, approximately 30 degrees C), high flow without added humidity (approximately 30 L/min), or a nebulizer attached to the tracheostomy tube without additional flow. The nebulizer was filled with 4 mL that contained 2.5 mg of albuterol, and operated at 8 L/min. The nebulizer was tested with a T-piece or tracheostomy mask. For the metered-dose inhaler experiments, a spacer was used and actuation of the inhaler (100 microg per actuation) was synchronized with inhalation (4 actuations separated by > or = 15 s). When the spacer was used without additional flow, a valved T-piece was used with a 1-way valve placed either proximal or distal to the spacer. A filter was attached between the lung model and the distal end of the tracheostomy tube. Albuterol washed from the filter was measured by ultraviolet spectrophotometry.ResultsFor the nebulizer, the most efficient delivery was with no flow other than that to power the nebulizer and with a T-piece (p < 0.001). The most efficient method for aerosol delivery was metered-dose inhaler with a valved T-piece and placement of the 1-way valve in the proximal position (p < 0.001). The effect of humidity was unclear from the results of this study.ConclusionsAlbuterol delivery via tracheostomy was affected by the delivery device (nebulizer vs inhaler), bias gas flow, and the patient interface.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…