    Comparison of different flow-reducing bag-valve ventilation devices regarding respiratory mechanics and gastric inflation in an unprotected airway model.

    • Fabienne C Rabus, Heinz-Theo Luebbers, Klaus W Graetz, and Till S Mutzbauer.
    • Department of Craniomaxillofacial Surgery, University of Zurich Dental Center, Plattenstrasse 15, CH-8032 Zurich, Switzerland.
    • Resuscitation. 2008 Aug 1;78(2):224-9.

    ObjectiveGastric inflation (GI) is a significant issue when ventilation is performed on unprotected airways.DesignExperimental analysis on the respiratory effects of hose extended bag-valve ventilation devices designed to reduce inspiratory pressure and flow.SettingLaboratory with lung/oesophageal sphincter simulator and pressure-flow-volume analyser. Lung compliance: 300ml/kPa, airway resistance: 0.5kPa/l/s. Lower oesophageal sphincter pressure (LOSP): 0.5kPa.InterventionsBag-valve ventilation of lung simulator. Twelve academic dental staff members used four devices: Ambu Mark III attached to either a reservoir bag (R) or a pressure relief valve (SV), SMART BAG (SB), and Easy Grip (EG) as control.ResultsAfter Bonferroni correction (p-level of significance 0.0083) for multiple comparisons, no evidence of difference between inspiratory tidal volumes (TVIN) administered by use of R (median 137ml) and SB (149ml) was found. Differences in TVIN were only detected between R and SV (188ml) (p=0.002). Only a trend towards TVIN differences between SB and R in comparison to EG (195ml) was found (p=0.009). Distributions of peak pressures differed when R (median 0.7kPa) and SV (1.0kPa) (p=0.006) or SB (0.7kPa) and SV (p=0.002) were compared. Peak inspiratory flow rates differed between EG (median 59l/min) and R (32l/min) as well as SB (42l/min) and between SB and SV (50l/min) (all with p=0.001). GI was lowest by use of R (median 103ml) compared to all other devices (EG: 518ml, SV: 394ml, SB: 271ml) (p=0.001). The areas under the pressure/flow over time curves were larger during SB compared to R ventilation. Mean airway pressures were significantly lower by use of R (0.1kPa) compared to SB (0.3kPa) (p<0.008).ConclusionLowering GI by pressure-flow reduction may result in lower TV depending on the device used. Lowest GI resulted from R ventilation. This may be explained by the specific pressure/time or flow/time patterns achieved by use of this device.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.