    A novel, versatile valved holding chamber for delivering inhaled medications to neonates and small children: laboratory simulation of delivery options.

    • Robert M DiBlasi, Dominic P Coppolo, Mark W Nagel, Cathy C Doyle, Valentina I Avvakoumova, Rubina S Ali, and Jolyon P Mitchell.
    • Center for Developmental Therapeutics, Seattle Children's Research Institute, Washington, USA.
    • Resp Care. 2010 Apr 1;55(4):419-26.

    BackgroundDelivery of bronchodilator to infants and small children from a pressurized metered-dose inhaler with valved holding chamber (pMDI-VHC) is limited by airway narrowness, short respiratory cycle time, and small tidal volume (V(T)). There is a need for a versatile, efficient VHC, given the variety of treatment modalities.MethodsWe tested the AeroChamber Mini VHC (the internal geometry of which is optimized for aerosol delivery, and which accepts a pMDI canister that has a dose counter) in experiments to determine differences in the delivery of hydrofluoroalkane-propelled albuterol (90 microg/actuation) during: mechanical ventilation via endotracheal tube (ETT); manual resuscitation via ETT; and spontaneous breathing via face mask. We tested 5 units of the AeroChamber Mini VHC per test. We simulated the tidal breathing of a premature neonate (V(T) 6 mL), a term neonate (V(T) 20 mL), and a child approximately 2 years old (V(T) 60 mL). We collected the aerosol on an electret filter and quantitatively assayed for albuterol.ResultsThe total emitted mass of albuterol per actuation that exited the VHC was marginally greater during spontaneous breathing (12.1 +/- 1.8 microg) than during manual resuscitation (10.0 +/- 1.1 microg) (P = .046). Albuterol delivery via mechanical ventilation, though comparable with the premature-neonate model (3.3 +/- 1.2 microg), the term-neonate model (3.8 +/- 2.1 microg), and the 2-y-old-child model (4.2 +/- 2.3 microg) (P = .63), was significantly lower than in the spontaneous-breathing and manual-resuscitation models (P < .001). In the neonatal models the total emitted mass was similar with the spontaneous-breathing model (6.0 +/- 1.0 microg with the premature-neonate model, 10.5 +/- 0.7 microg with the term-neonate model) and the manual-resuscitation model (5.5 +/- 0.3 microg premature-neonate model, 10.7 +/- 0.9 microg term-neonate model) (P > or = .46 via one-way analysis of variance).ConclusionThe reduced delivery of albuterol during mechanical ventilation (compared to during spontaneous breathing and manual resuscitation via ETT) was probably associated with the saturated atmosphere in the breathing circuit (37 degrees C, relative humidity > 99%), compared to the ambient air (22 +/- 1 degrees C, 44 +/- 7% relative humidity). The AeroChamber Mini VHC may provide a versatile alternative to VHCs that are designed exclusively for one aerosol treatment modality.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.