• J Aerosol Med Pulm Drug Deliv · Apr 2014

    Influence of exhalation valve and nebulizer position on albuterol delivery during noninvasive positive pressure ventilation.

    • Bing Dai, Jian Kang, Long-feng Sun, Wei Tan, and Hong-wen Zhao.
    • Department of Respiratory Medicine, First Affiliated Hospital of China Medical University , Shenyang, 110001, People's Republic of China .
    • J Aerosol Med Pulm Drug Deliv. 2014 Apr 1;27(2):125-32.

    BackgroundEarly studies have found better clinical efficiency when a nebulizer was used with noninvasive positive pressure ventilation (NPPV), compared with spontaneous breathing without NPPV. However, very limited research addressed factors that might affect aerosol delivery. This study aimed to investigate the influence of exhalation valves and nebulizer positions on aerosol delivery during NPPV.MethodsWe determined the efficiency of aerosol delivery in patients receiving NPPV with a lung model that simulates spontaneous breathing. Single-arch exhalation port, plateau exhalation valve, and whisper swivel were chosen as exhalation valves under different levels of inspiratory and expiratory pressures. A nebulizer was filled with 1 mL of 0.5% albuterol solution in 3 mL of normal saline, driven with 8 L/min oxygen, and placed at either a proximal position in the ventilator circuit (near the ventilator outlet, where humidifiers are usually connected) or a distal position in the ventilator circuit (between exhalation valve and lung model connection). Albuterol was collected by filters and then measured by ultraviolet spectrophotometry. The velocities of gas flow were also measured at different nebulizer positions.ResultsSignificant differences in the gas flow velocity were shown between proximal and distal positions of the breathing circuit under four combinations of inspiratory and expiratory pressure levels (15/5, 15/10, 25/5, and 25/10 cmH2O) (p<0.05). When the nebulizer was positioned distally, the single-arch exhalation port had the highest aerosol delivery, and the whisper swivel had the lowest aerosol delivery (p<0.05). When the nebulizer was placed proximally, the single-arch exhalation port had lower efficiency of aerosol delivery than the whisper swivel and plateau exhalation valve (p<0.05). In addition, higher inspiratory pressure was associated with increased aerosol delivery (p<0.05). The influence of expiratory pressure on aerosol delivery appeared too complex to predict.ConclusionsThe type of exhalation valve and the position of the nebulizer in the ventilator circuit have a significant influence on the efficiency of aerosol delivery during NPPV. As a result, with different exhalation valves, an appropriate nebulizer position should be carefully chosen, and the inhaled dose should be adjusted after accurate prediction of aerosol delivery to ensure optimal clinical efficacy.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…