    Laboratory and clinical evaluation of the MAX transport ventilator.

    • J A Johannigman, R D Branson, R Campbell, and J M Hurst.
    • Department of Surgery, Wilford Hall, United States Air Force, San Antonio, TX.
    • Resp Care. 1990 Oct 1;35(10):952-9.

    UnlabelledTransport of critically ill, mechanically ventilated patients from intensive care units for diagnostic and therapeutic procedures has become common in the last decade. Maintenance of adequate oxygenation and ventilation during transport is mandatory. We evaluated the Hamilton MAX transport ventilator in the laboratory and in the clinical arena to determine its usefulness during in-hospital transport.MethodsIn the laboratory, we determined the MAX's ability to assure tidal volume (VT) delivery in the face of decreasing compliance of a test lung, and we tested the alarm system. Using a two-compartment lung model modified to simulate spontaneous breathing, we also evaluated the responsiveness of the demand valve. The clinical evaluation was accomplished by comparing arterial blood gases and ventilator settings in the intensive care unit to those during transport.ResultsAs lung compliance was reduced from 0.1 to 0.02 L/cm H2O [1.0 to 0.20 L/kPa], delivered VT fell significantly at each set VT. The alarm systems performed according to manufacturer's specifications. The demand valve triggered appropriately without positive end-expiratory pressure (PEEP), but as PEEP was increased, triggering became more difficult. The demand valve is referenced to ambient pressure and cannot compensate for elevated end-expiratory pressures. During patient transport, arterial blood gases were comparable to those achieved in the ICU. Because an inspired oxygen concentration of 1.0 was used during transport, arterial oxygenation (PaO2) was significantly greater (123 +/- 75 vs 402 +/- 85 torr [16.4 +/- 10 vs 53.6 +/- 11 kPa]). A higher ventilator rate was required during transport to prevent tachypnea (7 +/- 3 vs 12 +/- 6 breaths/min), and peak inspiratory pressure (PIP) was higher during transport (40 +/- 8 vs 52 +/- 11 cm H2O [3.9 +/- 0.8 vs 5.1 +/- 1.1 kPa]).ConclusionsThe MAX is a reliable transport ventilator, capable of maintaining adequate ventilation and oxygenation in a majority of mechanically ventilated patients. Care should be taken to assure adequate VT delivery at high PIP, and ventilator rate may require adjustment to prevent tachypnea associated with triggering the non-PEEP-compensated demand valve when PEEP greater than 8 cm H2O [0.8 kPa] is used.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.