Detachment of the tip of a motorized shaver within the knee joint: a complication of arthroscopic surgery.
- Yong In, Won-Jong Bahk, and Jong-Beom Park.
- Department of Orthopaedic Surgery, Uijongbu St. Mary's Hospital, The Catholic University of Korea, Uijongbu, Kyunggi Province, South Korea. iy1000@cmc.cuk.ac.kr
- Arthroscopy. 2003 Jul 1;19(6):E25-7.
AbstractIn arthroscopic procedures, instrument failures are rare complications with low incidences. We present a case of detachment of the tip of the motorized shaver within the knee joint during an arthroscopic anterior cruciate ligament reconstruction procedure. This is the first report of its kind to our knowledge. We did not realize that the tip was missing within the knee joint during the surgery. We believe that the design of the shaver blade that was used needs modification, and surgeons must take special precautions to inspect the instrument before and after placing it in the joint.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.