• Anesthesia and analgesia · Apr 2012

    Comparative Study

    Poor accuracy of noninvasive cardiac output monitoring using bioimpedance cardiography [PhysioFlow(R)] compared to magnetic resonance imaging in pediatric patients.

    • Katherine Taylor, Cedric Manlhiot, Brian McCrindle, Lars Grosse-Wortmann, and Helen Holtby.
    • Department of Anesthesia, Hospital for Sick Children, 555 University Ave., Toronto, ON M5G 1X8, Canada. katherine.taylor@sickkids.ca
    • Anesth. Analg.. 2012 Apr 1;114(4):771-5.

    BackgroundIdentification of low cardiac output (CO) states in anesthesia is important because preoperative hemodynamic optimization may improve outcome in surgery. Accurate real-time CO measurement would be useful in optimizing "goal-directed" therapy. We sought to evaluate the reliability and accuracy of CO measurement using bioimpedance cardiography (PhysioFlow®, NeuMeDx, Bristol, PA) in pediatric patients with and without cardiac disease undergoing anesthesia for magnetic resonance imaging (MRI).MethodsAll consenting patients undergoing anesthesia for cardiac MRI were enrolled. After equilibration of anesthesia for ≥10 minutes, 6 PhysioFlow electrodes were applied to the patient's chest for continuous real-time monitoring for 10 minutes. Data were stored in 15-second epochs and later averaged offline to obtain CO. Phase contrast MRI measurements of flow volumes in the superior vena cava and ascending and descending aorta were made from a single imaging plane through all 3 vessels at the level of the right pulmonary artery. Both CO measurements were indexed to body surface area. The anesthetic technique was the same for both measurements. Agreement was assessed using Bland-Altman analysis.ResultsThirty-one patients were enrolled and 23 were analyzed. The median age at study was 2.8 years (range, 0.02-8.02 years) and median body surface area was 0.54 m(2) (range, 0.21-1.00 m(2)). Eleven of the 23 patients (48%) were males. Patients were grouped into those with univentricular physiology, 6 of 23 (26%); biventricular physiology with shunt, 3 of 23 (13%); biventricular without shunt, 10 of 23 (43%); and no structural heart disease, 4 of 23 (17%). The mean bias was -0.34 ± 1.50 L/min/m(2) (P = 0.29). The 95% limits of agreement were -3.21 to +2.69 L/min/m(2). Only 8 of 23 measurements (35%) were within 20% and 14 of 23 measurements (61%) were within 30% of each other.ConclusionPhysioFlow performance was not sufficiently accurate in this population. Modifications of the algorithm and further testing are required before this device can be recommended for routine clinical use in pediatric patients.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…