• Anaesthesia · Mar 1997

    Comparative Study Clinical Trial Controlled Clinical Trial

    Continuous versus intermittent thermodilution cardiac output measurement during orthotopic liver transplantation.

    • B W Böttiger, B Sinner, J Motsch, A Bach, H Bauer, and E Martin.
    • Department of Anaesthesia, University of Heidelberg, Germany.
    • Anaesthesia. 1997 Mar 1;52(3):207-14.

    AbstractWe evaluated intermittent and continuous thermodilution cardiac output data in 12 patients undergoing orthotopic liver transplantation. Measurements were performed at 16 predefined time points between induction of anaesthesia and 3 h after reperfusion of the liver graft. Cardiac output measurements yielded 192 data pairs (intermittent cardiac output range: 1.8-18.9 l.min-1, continuous cardiac output range: 3.3-20.0 l.min-1). During most of the procedure the correlation between intermittent and continuous cardiac output measurements was significant (r = 0.87, p < 0.0001), accompanied with a bias of -0.240 l.min-1 and a degree of precision of 1.789 l.min-1 (< 10.0 l.min-1:1.137 l.min-1, > or = 10.0 l.min-1:2.220 l.min-1). However, in the early phases after caval clamping and after reperfusion, accuracy was not acceptable. Only during these phases did the difference between the mean values of pulmonary artery blood temperature and rectal temperature increase (after caval clamping) or decrease (after reperfusion). In conclusion, despite acceptable levels of accuracy and precision between intermittent and continuous cardiac output measurement under stable conditions, both methods showed markedly decreased accuracy and precision in the early phases after caval clamping and after reperfusion, possibly owing to increased thermal noise.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…