• Resp Care · Jun 2002

    Imposed work of breathing during ventilator failure.

    • Paul N Austin, Robert S Campbell, Jay A Johannigman, and Richard D Branson.
    • Department of Nurse Anesthesia, Graduate School of Nursing, Uniformed Services University of the Health Sciences, Silver Spring, Maryland 20910, USA. paustin@usuhs.mil
    • Resp Care. 2002 Jun 1;47(6):667-74.

    IntroductionVentilators possess an anti-asphyxia valve that allows spontaneous breathing of ambient air during ventilator failure. This study examined the imposed work of breathing and pressure-time product of 8 critical care and 9 portable ventilators, using a laboratory simulation of spontaneous breathing during ventilator failure.MethodsA test lung was modified to simulate spontaneous breathing with a tidal volume of 0.5 L and peak inspiratory flow of 60 L/min. A pneumotachograph and pressure tap were placed at the proximal airway between the breathing circuit and endotracheal tube. Flow was derived from the pressure drop across the pneumotachograph. Signals were amplified, integrated, and saved to a spreadsheet program, and imposed work of breathing and pressure-time product were calculated. Also measured were the inspiratory pressure required to open the anti-asphyxia valve (cracking pressure), time to cracking pressure, maximum negative inspiratory pressure, and time to maximum negative inspiratory pressure.ResultsFor the critical care ventilators the mean +/- SD imposed work of breathing ranged from 213.07 +/- 3.53 to 890.63 +/- 0.88 mJ/L and the pressure-time product ranged from 2.67 +/- 0.01 to 13.37 +/- 0.01 cm H(2)O x s/L. For the portable ventilators the mean +/- SD imposed work of breathing ranged from 361.37 +/- 1.22 to 969.60 +/- 22.70 mJ/L and the pressure-time product ranged from 4.52 +/- 0.01 to 16.70 +/- 0.37 cm H(2)O x s/L.ConclusionsSpontaneous breathing during ventilator failure may impose work approximating the physiologic work of breathing. This imposed work may prevent effective breathing through the anti-asphyxia valve during mechanical ventilator failure due to electrical failure. These results reinforce the need to properly monitor mechanically ventilated patients and to have in place sufficient back-up power supplies and a method of manual ventilation.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.