• Pacing Clin Electrophysiol · Mar 2008

    Comparative Study

    Effects of surgical and endoscopic electrocautery on modern-day permanent pacemaker and implantable cardioverter-defibrillator systems.

    • Alan Cheng, Saman Nazarian, David D Spragg, Kenneth Bilchick, Harikrishna Tandri, Lynette Mark, Henry Halperin, Hugh Calkins, Ronald D Berger, and Charles A Henrikson.
    • Section for Cardiac Electrophysiology, Division of Cardiovascular Diseases, Department of Medicine, Johns Hopkins University School of Medicine, Baltimore, Maryland 21287, USA. alcheng@jhmi.edu
    • Pacing Clin Electrophysiol. 2008 Mar 1;31(3):344-50.

    BackgroundCurrent guidelines recommend that all implanted pacemakers (PPM) and defibrillators (ICD) be interrogated before and after every invasive procedure. The ability of newer devices to withstand system malfunction or failure during surgery/endoscopy remains unknown.ObjectiveTo determine the frequency of PPM or ICD malfunction from periprocedural electrocautery.MethodsNinety-two consecutive individuals referred for evaluation of a PPM or ICD system prior to noncardiac surgery/endoscopy were enrolled. Devices were preoperatively programmed to a "monitor only" zone to allow for detection of electromagnetic interferences (EMIs). Pacing parameters were maintained without disabling rate responsiveness. The devices were fully interrogated again after surgery. Correlations of inappropriate EMI sensing were made with reference to the distance from the site of electrocautery application to the device system.ResultsAll devices withstood periprocedural EMI exposure without malfunction or changes in programming. Minor changes in lead parameters were noted. Three device systems demonstrated brief atrial mode switching episodes, one of which was likely secondary to inappropriate sensing of atrial noise. Two pacemaker devices demonstrated inappropriate sensing of ventricular noise, both of which occurred when the application of electrocautery was within close proximity to the pacemaker generator (<8 cm). No ventricular sensed events were noted in any ICD system.ConclusionsEMIs during noncardiac surgical/endoscopic procedures pose little threat to current device systems. Rare occasions of inappropriate sensing by devices can be seen in situations where the application of unipolar electrocautery is in close proximity of the system. Routine postsurgical interrogation of PPM or ICDs may not be necessary.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.