• Resp Care · Jan 2003

    An in vitro study to investigate the use of a breath-actuated, small-volume, pneumatic nebulizer for the delivery of methacholine chloride bronchoprovocation agent.

    • Jolyon P Mitchell, Mark W Nagel, Sara Lou Bates, and Cathy C Doyle.
    • Trudell Medical International, London, Ontario, Canada. jmitchell@trudellmed.com.
    • Resp Care. 2003 Jan 1;48(1):46-51.

    BackgroundCurrent American Thoracic Society and American Association for Respiratory Care guidelines for the delivery of aerosol agents such as methacholine chloride (MC) for bronchoprovocation testing require the use of pneumatic jet nebulizers that have well-defined droplet size and mass output. A recently developed disposable, breath-actuated nebulizer (AeroEclipse) may offer bronchoprovocation testers an alternative to existing devices.MethodsWe studied the performance of 5 AeroEclipse nebulizers with regard to mass of MC delivered with various MC solution concentrations and numbers of inhalations, using a model of adult tidal breathing. Each nebulizer was operated with compressed air (8 L/min at 50 psig) and an initial fill of 2 mL. MC solutions with mass concentrations of 0.25, 0.98, 3.85, and 15.70 mg/mL were tested. The total mass of MC delivered was determined after 5, 10, and 15 complete breathing cycles, by assaying the MC collected on a filter placed at the nebulizer mouthpiece. The aerosol droplet size distribution, fine droplet fraction (FDF) (percentage of droplets < 4.8 microm diameter), and fine droplet mass (FDM) (mass of droplets < 4.8 microm diameter) were determined by laser diffractometry, using physiologically normal saline as a surrogate for MC solution.ResultsThe mean +/- SD FDM collected in 5 breathing cycles was 654 +/- 29 microg with the 15.70 mg/mL solution, 158 +/- 9 microg with the 3.85 mg/mL solution, 37 +/- 3 microg with the 0.98 mg/mL solution, and 7 +/- 2 microg with the 0.25 mg/mL solution. FDM showed a linear correlation (r(2) = 0.9999) with MC concentration, within the range studied. FDM also showed a linear correlation (r(2) = 0.999) with the number of breathing cycles. For instance, with the 15.70 mg/mL solution, FDM was 654 +/- 29 microg with 5 breathing cycles, 1,228 +/- 92 microg with 10 breathing cycles, and 1,876 +/- 132 microg with 15 breathing cycles.ConclusionsAlthough the bronchoprovocation test procedure had to be slightly modified from the guidelines to accommodate the operation of the AeroEclipse's breath-actuation feature, our measurements indicate that a predictable dose of MC, within the useful range for bronchoprovocation testing, can be delivered to an adult patient breathing tidally. The green indicator on the AeroEclipse could be used to coach the patient to inhale for a specific period, thereby controlling MC delivery per breathing cycle.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…