• Pacing Clin Electrophysiol · Jan 1999

    Multicenter Study Clinical Trial Controlled Clinical Trial

    Steroid eluting high impedance pacing leads decrease short and long-term current drain: results from a multicenter clinical trial. CapSure Z investigators.

    • K A Ellenbogen, M A Wood, D M Gilligan, M Zmijewski, and D Mans.
    • Department of Medicine, Medical College of Virginia, Richmond 23298-0053, USA. Kellenbogen@pol.net
    • Pacing Clin Electrophysiol. 1999 Jan 1;22(1 Pt 1):39-48.

    AbstractPacemaker lead technology has changed considerably over the past decades. The widespread use of low polarization highly porous electrodes and steroid elution electrodes has resulted in low chronic pacing thresholds, as well as a decrease in the incidence of exit block. Efforts to develop pacing leads with high impedance might theoretically lead to lower lead current drain, which is a component of battery capacity. Pulse generator longevity can be increased without sacrificing pacemaker capabilities if pacing current drain can be decreased. Decreasing the size of the stimulation electrode results in increased pacing impedance, and if pacing thresholds are unchanged, a decreased current drain is predicted by Ohm's law (I = V/R). There is limited data available on the pacing characteristics of large numbers of patients with high impedance leads, despite their recent general availability and increasing widespread use. This multicenter, controlled trial examined the differences in performance between standard steroid-eluting pacing leads in the atrium (Medtronic model 5524) and ventricle (Medtronic model 5024), and new high impedance steroid-eluting pacing leads in the atrium (Medtronic model 5534) and ventricle (Medtronic model 5034). Measurements of bipolar pacing thresholds at 2.5 V, pacing impedance, and sensing thresholds were determined within 24 hours of pacemaker implantation, and at 0.5, 1, 3, 6 and 12 months after pacemaker implantation in 609 patients. Pacing and sensing thresholds were similar for the control and high impedance leads at all times except for a slightly larger R wave with the high impedance leads at implantation and 12 months. The mean impedance of the high impedance pacing leads in the atrium and ventricle at 12 months was 992 +/- 175 and 1,080 +/- 220 omega, compared to 522 +/- 69 and 600 +/- 89 omega for the standard pacing leads in the atrium and ventricle (P < or = 0.001 for the high impedance leads compared to standard leads in each chamber). The mean atrial lead current (measured at 2.5 V) at 12 months was 2.6 +/- 0.5 mA with the high impedance lead, and 4.9 +/- 0.7 mA with the standard lead in the atrium (P < or = 0.001). In the ventricle, the mean lead current at 12 months was 2.4 +/- 0.4 mA with the high impedance pacing lead and 4.3 +/- 0.6 mA with the standard lead (P < or = 0.001). High impedance leads are associated with lower lead current drain than standard pacing leads in the atrium and ventricle for up to 1 year. No clinically important differences in sensing characteristics was noted with the high impedance leads in the atrium or ventricle compared to standard pacing leads. High impedance leads may result in increased pulse generator longevity.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…