Multicenter Study Clinical Trial Controlled Clinical TrialSteroid eluting high impedance pacing leads decrease short and long-term current drain: results from a multicenter clinical trial. CapSure Z investigators.
Pacemaker lead technology has changed considerably over the past decades. The widespread use of low polarization highly porous electrodes and steroid elution electrodes has resulted in low chronic pacing thresholds, as well as a decrease in the incidence of exit block. Efforts to develop pacing leads with high impedance might theoretically lead to lower lead current drain, which is a component of battery capacity. ⋯ High impedance leads are associated with lower lead current drain than standard pacing leads in the atrium and ventricle for up to 1 year. No clinically important differences in sensing characteristics was noted with the high impedance leads in the atrium or ventricle compared to standard pacing leads. High impedance leads may result in increased pulse generator longevity.