• J Clin Anesth · May 1997

    Randomized Controlled Trial Comparative Study Clinical Trial

    Effects of motion, ambient light, and hypoperfusion on pulse oximeter function.

    • N S Trivedi, A F Ghouri, N K Shah, E Lai, and S J Barker.
    • Department of Anesthesiology, University of California at Irvine, Orange 92668, USA.
    • J Clin Anesth. 1997 May 1;9(3):179-83.

    Study ObjectiveTo compare the performance of five pulse oximeters during hypoperfusion, probe motion, and exposure to ambient light interference.DesignProspective study.SettingLaboratory facility at a university medical center.Patients8 unanesthetized, ASA physical status I volunteers.InterventionsWe evaluated five common pulse oximeters with respect to three scenarios: (1) an operating room light was shone on oximeter probes, (2) a motion generator was used to generate 2 Hz and 4 Hz hand motion, and (3) a pneumatic compression device overlying the brachial artery was used to simulate hypoperfusion. Electrocardiographic (ECG) and arterial blood gas values were considered gold standards for heart rate (HR) and oxygen saturation (SpO2) respectively. SpO2 nondisplay and values greater than 4% from simultaneous arterial SaO2-oximeter values were defined as errors. Nondisplay of HR, or HR greater than 5% from ECG values, were also considered errors.Measurements And Main ResultsThe Ohmeda and Nellcor N200 with finger probe had the highest total failure rates with respect to both SpO2 and HR due to ambient light interference (p < 0.05). The Nellcor N200 with finger probe and N200 with C lock were the most accurate with regard to SpO2 during 2 Hz and 4 Hz motion (p < 0.05). However, all oximeters failed dramatically during 4 Hz motion when measuring HR. In the hypoperfusion model, the Nellcor N200 with finger probe and the Nellcor C Lock oximeters performed significantly better than all others in terms of both HR and SpO2 (P < 0.05), while the Criticare oximeter failed 100% of the time.ConclusionThere are significant differences in the accuracy of commercially available pulse oximeters during nonideal circumstances, with failure rates varying from approximately 5% to 50% depending on the oximeter and source of interference. Furthermore, no single oximeter performed the best under all conditions.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.