Ann Fr Anesth Reanim · Jan 1989
Case Reports[An unusual failure of the 900 C Siemens Servo ventilator].
- D Urli, B Calon, B Clever, and T Pottecher.
- Service d'anesthésie et de réanimation chirurgicale, Strasbourg.
- Ann Fr Anesth Reanim. 1989 Jan 1;8(4):369-70.
AbstractLung overinflation was observed in a patient ventilated by a Siemens Servo Ventilator 900 C. The expiratory valves failure to open was related to a transducer disconnection in the expiratory limb. This transducer controls opening of the expiratory valve and when disconnected expiratory valve remains closed.