• Resp Care · Feb 1982

    Oxygen delivery performance of four adult resuscitation bags.

    • T A Barnes and M E Watson.
    • Resp Care. 1982 Feb 1; 27 (2): 139-46.

    AbstractTo determine the fractional delivered oxygen concentrations (FDO2) and the proper functioning of patient valves in adult resuscitation bags, we studied the Laerdal, the PMR 1, the PMR 2, and the Vitalograph resuscitators at various oxygen flowrates and ventilation patterns, with and without reservoirs. The Vitalograph, with new or old reservoir, and the PMR 1 cannot deliver high oxygen concentrations (FDO2 greater than 0.90) and should not be used when high FDO2 is important. The PMR 2 and the Laerdal cannot deliver high oxygen concentrations without reservoirs attached; with reservoirs attached the Laerdal achieves FDO2 greater than 0.98 at a flowrate of 10 l/min, and the PMR 2 achieves FDO2 greater than 0.90 at a flowrate of 15 l/min.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.