• J Clin Anesth · May 1991

    Case Reports

    Bilateral tension pneumothorax due to a defective anesthesia breathing circuit filter.

    • C E Smith, J R Otworth, and P Kaluszyk.
    • Department of Anesthesiology, MetroHealth Medical Center, Cleveland, OH 44109.
    • J Clin Anesth. 1991 May 1; 3 (3): 229-34.

    AbstractA 25-year-old healthy patient developed bilateral tension pneumothorax during general endotracheal anesthesia due to a defective anesthesia breathing circuit filter. Prompt diagnosis and treatment of the pneumothorax was facilitated by the use of capnography and pulse oximetry. The manufacturing process of the breathing circuit and filter that made this accident possible has since been corrected by the manufacturer. Anesthesiologists must be alert to the possibility of such accidents with any breathing system using bacterial filters.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.