• Respiratory care · Mar 2015

    Performance of Ventilators Compatible With Magnetic Resonance Imaging: A Bench Study.

    • Yusuke Chikata, Nao Okuda, Masayo Izawa, Mutsuo Onodera, and Masaji Nishimura.
    • Medical Equipment Center, Tokushima University Hospital, Tokushima, Japan.
    • Respir Care. 2015 Mar 1;60(3):341-6.

    BackgroundMagnetic resonance imaging (MRI) is indispensable for diagnosing brain and spinal cord abnormalities. Magnetic components cannot be used during MRI procedures; therefore, patient support equipment must use MRI-compatible materials. However, little is known of the performance of MRI-compatible ventilators.MethodsAt commonly used settings, we tested the delivered tidal volume (V(T)), F(IO2), PEEP, and operation of the high-inspiratory-pressure-relief valves of 4 portable MRI-compatible ventilators (Pneupac VR1, ParaPAC 200DMRI, CAREvent MRI, iVent201) and one ICU ventilator (Servo-i). Each ventilator was set in volume control/continuous mandatory ventilation mode. Breathing frequency and V(T) were tested at 10 breaths/min and 300, 500, and 700 mL, respectively. The Pneupac VR1 has fixed V(T) and frequency combinations, so it was tested at V(T) = 300 mL and 20 breaths/min, V(T) = 500 mL and 12 breaths/min, and V(T) = 800 mL and 10 breaths/min. F(IO2) was 0.6 and 1.0. At the air-mix setting, F(IO2) was fixed at 0.5 with the Pneupac VR1, 0.45 with the ParaPAC 200DMRI, and 0.6 with the CAREvent MRI. PEEP was set at 5 and 10 cm H2O, and pressure relief was set at 30 and 40 cm H2O.ResultsV(T) error varied widely among ventilators (-28.1 to 25.5%). As V(T) increased, error decreased with the Pneupac VR1, ParaPAC 200DMRI, and CAREvent MRI (P < .05). F(IO2) error ranged from -13.3 to 25.3% at 0.6 (or air mix). PEEP error varied among ventilators (-29.2 to 42.5%). Only the Servo-i maintained V(T), F(IO2), and PEEP at set levels. The pressure-relief valves worked in all ventilators.ConclusionsNone of the MRI-compatible ventilators maintained V(T), F(IO2), and PEEP at set levels. Vital signs of patients with unstable respiratory mechanics should be monitored during transport and MRI.Copyright © 2015 by Daedalus Enterprises.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.