• Pediatrics · Mar 1986

    Comparative Study

    Limitations of self-inflating resuscitators.

    • N N Finer, K J Barrington, F Al-Fadley, and K L Peters.
    • Pediatrics. 1986 Mar 1; 77 (3): 417-20.

    AbstractIn an effort to characterize the performance of self-inflating resuscitators, three examples of three models were subjected to laboratory testing: the Ohio Hope II resuscitator (Ohio Medical Products, Madison, WI), the PMR-2 resuscitator (Puritan Medical Products, Lenexa, KS), and the Laerdal resuscitator (AS Laerdal, Stavanger, Norway). The devices were connected to a test lung and compressed at frequencies of from 10 to 60 and at greater than 60 breaths per minute at 5, 10, and 15 L/min of flow. These devices were used with and without a reservoir and were compressed at less than and more than the pop-off valve pressures. The results revealed that all devices equipped with reservoirs delivered a significantly higher oxygen concentration than those without (P less than .001), and that activation of the pop-off valve significantly reduced Fio2 under all conditions (P less than .005). The pop-off valves for each device were activated throughout a wide range of pressures, the Laerdal 41 to 72 cm H2O, PMR-2 51 to 97 cm H2O, and the Ohio Hope II 38 to 106 cm H2O, well in excess of the manufacturer's specifications. Only the Laerdal with reservoir was able to deliver an Fio2 of greater than 0.9 when compressed at more than the pop-off valve pressure at rates of up to 30 breaths per minute using flows of 10 L/min, and it was the only device to produce Fio2 values of greater than 0.9 at all rates to 60 breaths per minute when compressed at less than the pop-off valve pressure.(ABSTRACT TRUNCATED AT 250 WORDS)

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…