Paediatric anaesthesia · Aug 2015
Continuous minimally invasive cardiac output monitoring with the COstatus in a neonatal swine model: recalibration is necessary during vasoconstriction and vasodilation.
- Alvaro F Cisternas, Manuel Martin-Flores, and Robin D Gleed.
- Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY, USA.
- Paediatr Anaesth. 2015 Aug 1;25(8):852-9.
BackgroundThe COstatus monitor measures cardiac output via the transpulmonary ultrasound dilution method (COTPUD ) after injection of normal saline, and can calculate continuous cardiac output (CCO) from the arterial pressure waveform. The relationship between arterial waveform and COTPUD however, might be degraded during vasoconstriction/vasodilation.ObjectivesTo examine if recalibration of arterial waveform-derived CCO is required during mild vasoconstriction/vasodilation.MethodsIn 10 anesthetized piglets (6.6-10.1 kg), two COstatus monitors calculated the CCO from the same femoral arterial waveform before and during infusions of phenylephrine (PE; 1 or 3 mg·kg(-1) ·min(-1) ) and sodium nitroprusside (SNP; 1 or 5 mg·kg(-1) ·min(-1) ), administered in random order. One monitor was recalibrated (CCORecal ) after each intervention, while the other monitor was not (CCONon-Recal ). Recalibration was performed with COTPUD with 1 ml·kg(-1) normal saline as indicator. The effects of each infusion on hemodynamic parameters were compared with baseline using paired t-tests. The bias, limits of agreement (LOA), and percentage error between simultaneous measurements (CCORecal and CCONon-Recal ) were examined with Bland-Altman plots.ResultsInfusion of PE significantly increased COTPUD , heart rate (HR), and arterial pressures but not systemic vascular resistance (SVR). Infusion of SNP decreased arterial pressures without affecting COTPUD , HR, and SVR. There was no bias between CCORecal and CCONon-Recal at the baseline, but a small bias was observed during PE and SNP infusions. The LOA increased approximately 10 fold during vasoconstriction and vasodilation. The percentage error increased from ≤ 5% to 32% and 27% during PE and SNP infusions, respectively.ConclusionContinuous cardiac output (CO) measured with the COstatus monitor requires recalibration during vasoconstriction and vasodilation, even if changes in COTPUD or SVR are not substantial.© 2015 John Wiley & Sons Ltd.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.