Comparative Study
Albuterol aerosol delivered via metered-dose inhaler to intubated pediatric models of 3 ages, with 4 spacer designs.
- Piush Mandhane, Peter Zuberbuhler, Carlos F Lange, and Warren H Finlay.
- Department of Respirtatory Medicine, Hospital for Sick Children, Toronto, Ontario, Canada.
- Resp Care. 2003 Oct 1; 48 (10): 948-55.
ObjectiveTo determine the amount of albuterol, in various particle size ranges, delivered from a hydrofluoroalkane-propelled metered-dose inhaler (Airomir) in 3 models of pediatric intubation (ages 8 months, 4 years, and 16 years) using 4 types of aerosol reservoir: 3 spacers (ACE, AeroChamber HC MV, metal NebuChamber without 1-way valve) and 1 holding chamber (metal NebuChamber with 1-way valve).MethodsFive reservoirs of each type were tested with albuterol sulfate delivered via metered-dose inhaler that delivers 100 microg of albuterol per actuation. Each reservoir was connected to an endotracheal tube (ETT) that corresponded to the given patient age (8 months = 4 French; 4 years = 5 French; 16 years = 7.5 French) and to a valved system that allowed connection of the ETT to a cascade impactor. Simulated tidal volumes representative of children of the given ages were passed through the reservoir. Both the cascade impactor and the ETT were enclosed within a 100% humidity, 37 degrees C environment.ResultsFor the total amount of albuterol inhaled onto the impactor, and both the 1.1-4.7 microm and 1.1-3.3 microm inhaled fine-particle fractions, the NebuChamber-with-valve showed significantly greater drug delivery than the NebuChamber-without-valve, the AeroChamber HC MV, or the ACE (p < 0.001). Among the reservoirs without valves the NebuChamber showed significantly greater delivery than the AeroChamber HC MV or ACE (p < 0.001) for total drug deposition and for both the 1.1-4.7 microm and 1.1-3.3 microm fine-particle fractions. These results were consistent over all age groups. The AeroChamber HC MV had significantly greater delivery (total deposition) than the ACE (p < 0.001), except in the 4-year-old model. There were no significant differences between the AeroChamber HC MV and the ACE for either the 1.1-4.7 microm or the 1.1-3.3 microm fine-particle fraction.ConclusionAn aerosol reservoir with 1-way valve positioned between the spacer and the ETT improved the amount of inhaled albuterol 300-900%, compared to the other reservoirs.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.