• Der Anaesthesist · Dec 2009

    Case Reports

    [Intratracheal foreign body caused by a sheared endotracheal intubation stylet].

    • L Schaffranietz, F Graz, and A Tamke.
    • Klinik für Anästhesiologie und Intensivmedizin, KKH Delitzsch GmbH, 04509 Delitzsch. l.schaffranietz@kkh-delitzsch-gmbh.de
    • Anaesthesist. 2009 Dec 1; 58 (12): 1223-5.

    AbstractAfter orotracheal intubation of a 30-year-old man with a flexible tube, a floating foreign body was noticed inside the tube. The subsequent bronchoscopy revealed a second foreign body near the bifurcation of the trachea. Both foreign bodies were removed successfully and identified as parts of the plastic-sheathed stylet. Fatigue of material is discussed as the reason for the broken reusable stylet during intubation.