• Der Anaesthesist · Dec 2009

    Case Reports

    [Intratracheal foreign body caused by a sheared endotracheal intubation stylet].

    • L Schaffranietz, F Graz, and A Tamke.
    • Klinik für Anästhesiologie und Intensivmedizin, KKH Delitzsch GmbH, 04509 Delitzsch. l.schaffranietz@kkh-delitzsch-gmbh.de
    • Anaesthesist. 2009 Dec 1; 58 (12): 1223-5.

    AbstractAfter orotracheal intubation of a 30-year-old man with a flexible tube, a floating foreign body was noticed inside the tube. The subsequent bronchoscopy revealed a second foreign body near the bifurcation of the trachea. Both foreign bodies were removed successfully and identified as parts of the plastic-sheathed stylet. Fatigue of material is discussed as the reason for the broken reusable stylet during intubation.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.