• Foot Ankle Int · Nov 2013

    Comparative Study

    Biomechanical comparison of 4 different lateral plate constructs for distal fibula fractures.

    • Tobin T Eckel, Richard R Glisson, Prashanth Anand, and Selene G Parekh.
    • Department of Orthopaedic Surgery, Duke University Medical Center, Durham, NC, USA.
    • Foot Ankle Int. 2013 Nov 1; 34 (11): 1588-95.

    BackgroundDisplaced lateral malleolar fractures are often treated with reduction and surgical stabilization. However, there has not been a comprehensive laboratory comparison to determine the most appropriate device for treating these patients. This study subjected a range of contemporary lateral fibular plates to a series of mechanical tests designed to reveal performance differences.MethodsForty fresh frozen lower extremities were divided into 4 groups. A Weber B distal fibula fracture was simulated with an osteotomy and stabilized using 1 of 4 plate systems: a standard Synthes one-third tubular plate with interfragmentary lag screw, a Synthes LCP locking plate with lag screw, an Orthohelix MaxLock Extreme low-profile locking plate with lag screw, or a TriMed Sidewinder nonlocking plate. Controlled monotonic bending and cyclic torsional loading were applied and bending stiffness, torsional stiffness, and fracture site motion were quantified. Resistance to cyclic torsional loading was determined by quantifying the number of loads withstood before excessive rotation occurred. Correlation between bone mineral density and each of the mechanical measures was determined.ResultsThere was no difference in angulation or bending stiffness between plates. All plates except the LCP showed greater lateral deflection than in the other bending directions. Bending stiffness was lowest in lateral distal fragment deflection for all 4 plates. There was a positive correlation between bone mineral density and bending stiffness for all plate types. There was no difference in fracture site rotation between plate types in internal or external torsion, but internal rotation of the distal fragment consistently exceeded external rotation. Torsional stiffness in external rotation exceeded stiffness in internal rotation in nearly all specimens. LCP plates performed relatively poorly under cyclic torsion.ConclusionsSignificant differences in plate performance were not demonstrated. The effects of bone quality variability and differences in interfragmentary screw purchase resulted in data dispersion that confounded absolute ranking of plate performance.Clinical RelevanceIdentification of an optimal lateral fibular plating system has the potential to improve the clinical outcome of malleolar fracture fixation, particularly when patient conditions are unfavorable.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.