• J Anal Toxicol · Oct 1998

    Multicenter Study Comparative Study Clinical Trial

    A multiple-site laboratory evaluation of three on-site urinalysis drug-testing devices.

    • D J Crouch, J F Frank, L J Farrell, H M Karsch, and J E Klaunig.
    • University of Utah and Center for Human Toxicology, Salt Lake City 84112, USA.
    • J Anal Toxicol. 1998 Oct 1; 22 (6): 493-502.

    AbstractPresented are findings from a multisite laboratory evaluation comparing on-site urinalysis drug-test results to results from Syva EMIT immunoassay and gas chromatography-mass spectrometry (GC-MS). Three laboratories participated in the NHTSA-funded project. Specimens were tested for amphetamines, benzodiazepines, cocaine, cannabinoids, and opiates. Each laboratory selected 20 urines that tested positive for a single drug/drug class and 20 that tested negative to challenge the on-site drug-testing devices. Qualitative and quantitative GC-MS confirmations were performed to ensure that all positive samples contained the target drug(s)/metabolite(s) and that all negative samples did not contain the target analytes. EZ-SCREEN, ONTRAK, and TRIAGE on-site test kits were selected for evaluation. On-site false-positive results, in which GC-MS-verified negative urine samples gave positive on-site results, were rare. Two such errors were recorded with both EZ-SCREEN and TRIAGE. Cross-reactivity from samples containing GC-MS-verified high concentrations of alternate drugs was also rare. One cross-reactive error was recorded while testing for cocaine with EZ-SCREEN, a second while testing for benzodiazepines with ONTRAK, and a third while testing for cocaine with ONTRAK. The EZ-SCREEN kit did not appear to adhere to a cutoff concentration as demonstrated by the number of samples that contained low concentrations of the target drugs that tested positive with this device. A significant finding of this study was that comparing on-site test device results with those of EMIT for samples with drug concentrations near the reporting cutoff was very complex. It required a thorough knowledge of the performance of each device, EMIT, and GC-MS. It also required an investigation of each discrepant result-a consideration not taken in many previous evaluations of on-site testing devices. Compared with current federal guidelines for workplace urinalysis testing, more donor samples would screen positive for cannabinoids and cocaine by the on-site devices than by EMIT immunoassay. However, fewer would be reported as positive because most contained GC-MS-determined drug concentrations lower than the federal confirmation and reporting limits.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

What will the 'Medical Journal of You' look like?

Start your free 21 day trial now.

We guarantee your privacy. Your email address will not be shared.