• J Clin Monit Comput · Feb 2018

    Feasibility of automated titration of vasopressor infusions using a novel closed-loop controller.

    • Joseph Rinehart, Michael Ma, Michael-David Calderon, and Maxime Cannesson.
    • Department of Anesthesiology & Perioperative Care, University of California Irvine, 101 The City Dr. South, Orange, CA, 92868, USA. jrinehar@uci.edu.
    • J Clin Monit Comput. 2018 Feb 1; 32 (1): 5-11.

    AbstractBlood pressure management is a central concern in critical care patients. For a variety of reasons, titration of vasopressor infusions may be an ideal use-case for computer assistance. Using our previous experience gained in the bench-to-bedside development of a computer-assisted fluid management system, we have developed a novel controller for this purpose. The aim of this preliminary study was to assess the feasibility of using this controller in simulated patients to maintain a target blood pressure in both stable and variable blood-pressure scenarios. We tested the controller in two sets of simulation scenarios: one with stable underlying blood pressure and a second with variable underlying blood pressure. In addition, in the variable phase of the study, we tested infusion-line delays of 8-60 s. The primary outcome for both testing conditions (stable and variable) was % case time in target range. We determined a priori that acceptable performance on the first phase of the protocol would require greater than 95% case-time in-target given the simple nature of the protocol, and for the second phase of the study 80% or greater given the erratic nature of the blood pressure changes taking place. 250 distinct cases for each simulation condition, both managed and unmanaged, were run over 4 days. In the stable hemodynamic conditions, the unmanaged group had an MAP of 57.5 ± 4.6 mmHg and spent only 5.6% of case time in-target. The managed group had an MAP of 70.3 ± 2.6 and spent a total of 99.5% of case time in-target (p < 0.00001 for both comparisons between groups). In the variable hemodynamic conditions, the unmanaged group had an MAP of 53.1 ± 5.0 mmHg and spent 0% of case time in-target. The managed group had an MAP of 70.5 ± 3.2 mmHg (p < 0.00001 compared to unmanaged group) and spent 88.6% of case time in-target (p < 0.00001 compared to unmanaged group), with 6.4% of case time over and 5.1% of case time under target. Increasing infusion lag increased coefficient of variation by about 10% per 15 s of lag (p = 0.001). This study demonstrated that this novel controller for vasopressor administration is able to main a target mean arterial pressure in a simulated physiologic model in the face of random disturbances and infusion-line lag.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.