Review Case Reports
Balloon Angioplasty for in-stent restenosis resulting in carotid stent fracture with literature review of stent-in-stent technique as a viable therapeutic option.
- Abdul-Haseeb Naeem, Ghaya Al-Rumaihi, Amirpouyan Namavarian, Manas Sharma, and Melfort Boulton.
- Division of Neurosurgery, Department of Clinical Neurological Sciences, London Health Sciences Centre, London, Ontario, Canada. Electronic address: abdul.naeem@lhsc.on.ca.
- World Neurosurg. 2017 May 1; 101: 818.e1-818.e6.
BackgroundCarotid stent fractures are rare, and multiple etiologies have been proposed to explain their occurrence. We describe a patient with an internal carotid artery (ICA) stent who developed in-stent restenosis. We performed balloon angioplasty to address in-stent restenosis, but he developed a carotid stent fracture after the procedure. To our knowledge, balloon angioplasty has not been reported to cause stent fractures.Case DescriptionA 72-year-old man underwent stent placement for symptomatic left ICA stenosis with residual stenosis of 55% after stent placement. He presented with transient ischemic attacks 2 months later, and work-up revealed in-stent restenosis of the left ICA. Given prior complete occlusion of right ICA and right vertebral artery and narrowing of left vertebral artery ostium, satisfactory balloon (5 × 40 mm) angioplasty was carried out. After balloon angioplasty, x-ray showed a new stent fracture, which was initially missed on immediate postoperative imaging. He presented 9 months later with symptoms of compromised cerebral perfusion. Work-up revealed the previously missed stent fracture causing blood flow changes. Peak systolic velocity in the left ICA was 383 cm/second. He underwent left ICA repeat stent placement via a stent-in-stent technique for symptomatic severe left ICA stenosis of 70% with 40% residual stenosis after new stent deployment.ConclusionsBalloon angioplasty to address in-stent restenosis can secondarily cause stent fractures. We provide evidence of successful management of stent fracture with recurrent in-stent stenosis by repeat stent placement via a stent-in-stent technique.Copyright © 2017 Elsevier Inc. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.