• Resp Care · Apr 2009

    Optimum design parameters for a therapist-constructed positive-expiratory-pressure therapy bottle device.

    • Régis Gemerasca Mestriner, Rafael Oliveira Fernandes, Luís Carlos Steffen, and Márcio Vinícius Fagundes Donadio.
    • Faculdade de Enfermagem, Nutrição e Fisioterapia, Pontifícia Universidade Católica do Rio Grande do Sul, Av Ipiranga, 6681, Prédio 12-8 degrees Andar, Partenon, Porto Alegre, Rio Grande do Sul, Brasil.
    • Resp Care. 2009 Apr 1; 54 (4): 504-8.

    BackgroundPositive-expiratory-pressure (PEP) therapy uses positive airway pressure generated by a either a fixed-orifice resistor or a threshold resistor. We hypothesized that tubing diameter and length, and the diameter of the PEP bottle's air-escape orifice would impact the PEP pressure delivered to the airway and determine whether the PEP bottle acts as a threshold resistor or a fixed-orifice resistor.MethodsWe designed a model composed of a bottle partially filled with water, a compressed air source, a pneumotachometer, and a manometer, to evaluate the effects of various tubing diameters (range 2-25 mm inner diameter) and lengths (range 20-80 cm long). In the first set of experiments, the PEP bottle had an open top, so there was no pressure other than the atmospheric pressure against the air escaping from the immersed tubing. The distal tip of the tube was 10 cm below the surface of the water (ie, a water-column pressure of 10 cm H(2)O), and we tested flows of 1, 5, 10, 15, 20, and 25 L/min. In the second set of experiments we tested a PEP bottle, the top of which was closed except for an air-escape orifice (4, 6, 8, 9, or 10 mm).ResultsWith tubing of 2-6 mm inner diameter, the length of the tubing and the flow significantly affected the PEP pressure (ie, the system was not a threshold resistor). With tubing > or = 8 mm inner diameter there were no significant PEP-pressure differences with any of the tubing lengths or flows tested, which indicates a threshold-resistor system. The 4-mm and 6-mm air-escape orifices significantly increased the PEP pressure, whereas the 8 mm air-escape orifice did not increase the PEP pressure.ConclusionsTo obtain a threshold-resistor PEP bottle system (ie, the PEP pressure is generated only by the water-column pressure), the tubing must be > or = 8 mm inner diameter, and the air-escape orifice must be > or = 8 mm.

      Pubmed     Free full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…