• Med Biol Eng Comput · May 2001

    Error analysis of a natural breathing calibration method for respiratory inductive plethysmography.

    • N O Strömberg.
    • Department of Biomedical Engineering, Linköping University, Sweden. tomst@imt.liu.se
    • Med Biol Eng Comput. 2001 May 1; 39 (3): 310-4.

    AbstractRespiratory volumes are measured non-invasively by recording rib cage and abdominal motions using respiratory inductive plethysmography (RIP). Qualitative diagnostic calibration (QDC) of RIP is based on the natural variability in the relative rib-cage-to-abdomen contribution during tidal breathing. ODC does not require subject cooperation but it has previously been shown that accuracy may deteriorate when breathing pattern changes. The aim of this study was to investigate the causes and situations where QDC accuracy deteriorates. The QDC method was compared to PRA (calibration during voluntarily preferential rib cage or abdomen breathing) in ten adults. A reference RIP calibration was obtained from all validation data (REF). The PRA method had better accuracy than the ODC method (p<0.01). The volumetric error ranged between 10% and 136% with QDC and between 5% and 33% with PRA. The PRA calibration factors were within 6% of those from REF, while the QDC rib-cage factor was underestimated by 15% and the abdominal factor was overestimated by 38%. Small natural variability in the relative rib-cage-to-abdomen contribution was related to poor accuracy. Each compartment's variability depended on its magnitude, which is a violation of the QDC assumptions.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…