Multicenter Study Clinical Trial
Delirium detection using relative delta power based on 1-minute single-channel EEG: a multicentre study.
- T Numan, M van den Boogaard, A M Kamper, P J T Rood, L M Peelen, Slooter A J C AJC Department of Intensive Care Medicine and Brain Center Rudolf Magnus, University Medical Center Utrecht, Utrecht University, Utrecht, The Netherlands., and Dutch Delirium Detection Study Group.
- Department of Intensive Care Medicine and Brain Center Rudolf Magnus, University Medical Center Utrecht, Utrecht University, Utrecht, The Netherlands. Electronic address: tiannenuman@gmail.com.
- Br J Anaesth. 2019 Jan 1; 122 (1): 60-68.
BackgroundDelirium is frequently unrecognised. EEG shows slower frequencies (i.e. below 4 Hz) during delirium, which might be useful in improving delirium recognition. We studied the discriminative performance of a brief single-channel EEG recording for delirium detection in an independent cohort of patients.MethodsIn this prospective, multicentre study, postoperative patients aged ≥60 yr were included (n=159). Before operation and during the first 3 postoperative days, patients underwent a 5-min EEG recording, followed by a video-recorded standardised cognitive assessment. Two or, in case of disagreement, three delirium experts classified each postoperative day based on the video and chart review. Relative delta power (1-4 Hz) was based on 1-min artifact-free EEG. The diagnostic value of the relative delta power was evaluated by the area under the receiver operating characteristic curve (AUROC), using the expert classification as the gold standard.ResultsExperts classified 84 (23.3%) postoperative days as either delirium or possible delirium, and 276 (76.7%) non-delirium days. The AUROC of the relative EEG delta power was 0.75 [95% confidence interval (CI) 0.69-0.82]. Exploratory analysis showed that relative power from 1 to 6 Hz had significantly higher AUROC (0.78, 95% CI 0.72-0.84, P=0.014).ConclusionsDelirium/possible delirium can be detected in older postoperative patients based on a single-channel EEG recording that can be automatically analysed. This objective detection method with a continuous scale instead of a dichotomised outcome is a promising approach for routine detection of delirium.Clinical Trial RegistrationNCT02404181.Copyright © 2018 British Journal of Anaesthesia. Published by Elsevier Ltd. All rights reserved.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.