• Lasers Surg Med · Jan 2003

    Temperature controlled burn generation system based on a CO2 laser and a silver halide fiber optic radiometer.

    • Meir Cohen, Avi Ravid, Vered Scharf, Daniel Hauben, and Abraham Katzir.
    • Schneider Children's Medical Center of Israel, Department of Plastic Surgery and Burn Unit, Rabin Medical Center, Beilinson Campus, Tel-Aviv University, Tel-Aviv, Israel. drcohen@netvision.net.il
    • Lasers Surg Med. 2003 Jan 1; 32 (5): 413-6.

    Background And ObjectivesExperimental animal study of burns is dependent on a reliable burn generation system. Most of the experimental systems used today are unable to produce precise partial thickness burns. This limits the ability to study minor changes associated with burn care. The aim of the study was to develop a method for generating burns with a fixed depth using a CO2 laser burn generation system.Materials And MethodsThe burn generation system was composed of two components: a burn generation device and a temperature sensing and control system. These components were designed to operate together in order to keep a constant, predetermined skin surface temperature during prolonged burn generation. One hundred thirty-eight spot burns were generated on the back of five shaved 450 g male Wistar rats. The rat skin was exposed to a 70 degrees C for 5-60 seconds. The burned areas were excised and underwent evaluation by hematoxylin-eosin-stained slide microscopy.ResultsA linear correlation was found between the duration of exposure and the average burn depth (r = 0.93). This correlation is represented by the equation: burn depth in millimeters = 0.012x (duration in seconds of skin exposure at 70 degrees C).ConclusionsThe fiber-optic-controlled laser burn generation system studied is a reliable tool for creating partial thickness as well as full thickness skin burns in rats.Copyright 2003 Wiley-Liss, Inc.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.