Comparative StudyMeasurement of heparin concentration in whole blood with the Hepcon/HMS device does not agree with laboratory determination of plasma heparin concentration using a chromogenic substrate for activated factor X.
- J F Hardy, S Bélisle, D Robitaille, J Perrault, M Roy, and L Gagnon.
- Department of Anesthesia, Montreal Heart Institute, Quebec, Canada.
- J. Thorac. Cardiovasc. Surg. 1996 Jul 1; 112 (1): 154-61.
UnlabelledMeasurement of circulating heparin concentration has been suggested to optimize anticoagulation during cardiopulmonary bypass. The Hepcon/HMS device (Medtronic HemoTec, Inc., Parker, Colo.) uses heparin/protamine titration to quantitatively determine heparin concentration. Extensive validation of this instrument is still lacking.MethodsAgreement between heparin concentrations measured by the Hepcon/HMS system and by laboratory determination was evaluated in 16 patients undergoing cardiac operations. For laboratory determinations, plasma heparin concentration was derived from the measure of anti-Xa activity by means of chromogenic substrate technique. The Hepcon/HMS instrument and cartridges measured whole blood heparin concentration. Samples were analyzed 5 minutes after administration of heparin, 15 and 30 minutes after the start of cardiopulmonary bypass, 5 minutes after aortic unclamping, at the end of cardiopulmonary bypass, and after administration of protamine. Data were plotted and interpreted according to the method of Bland and Altman: First, a difference less than 1.4 U/ml (i.e., +/- 0.7 U/ml) was chosen as acceptable, because it would not cause major difficulties in clinical interpretation; second, the difference between the two measurement techniques was plotted against the mean of the two measures.ResultsThe mean difference (bias) between heparin concentrations derived by the Hepcon/HMS device and those obtained by laboratory determination was as expected for measures performed on whole blood versus plasma (1.45 U/ml). Nevertheless, heparin concentrations derived by the Hepcon/HMS device may be as much as 2.76 U/ml above or 6.17 U/ml below the concentrations measured in the laboratory, differences well outside the predetermined limits of agreement and clearly unacceptable for clinical purposes.ConclusionWe conclude that heparin concentrations determined with the Hepcon/HMS instrument do not agree with laboratory determination of heparin concentration. Monitoring of heparin concentrations during bypass with the Hepcon/HMS device cannot be recommended.
Notes
Knowledge, pearl, summary or comment to share?You can also include formatting, links, images and footnotes in your notes
- Simple formatting can be added to notes, such as
*italics*
,_underline_
or**bold**
. - Superscript can be denoted by
<sup>text</sup>
and subscript<sub>text</sub>
. - Numbered or bulleted lists can be created using either numbered lines
1. 2. 3.
, hyphens-
or asterisks*
. - Links can be included with:
[my link to pubmed](http://pubmed.com)
- Images can be included with:
![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
- For footnotes use
[^1](This is a footnote.)
inline. - Or use an inline reference
[^1]
to refer to a longer footnote elseweher in the document[^1]: This is a long footnote.
.