• J Clin Monit Comput · Apr 2002

    Improving accuracy of the chemiluminescent analyser for measurement of nitric oxide and nitrogen dioxide in respired gas.

    • Robert F Salamonsen.
    • Alfred Hospital, Melbourne, Australia. r.salamonsen@alfred.org.au
    • J Clin Monit Comput. 2002 Apr 1; 17 (3-4): 173-9.

    ObjectiveThe aim of this study was to improve the accuracy of the chemiluminescent analyser in the measurement of nitric oxide (NO) and nitrogen dioxide (NO2) in both dry and humidified gas mixtures containing oxygen (O2) in varying concentrations. This work was performed because initial attempts to determine rate constants for the decay of NO and rates of formation of NO2 were unsuccessful.MethodsEquations were developed by which the raw output signal of the analyser could be corrected for errors arising from variations in sensitivity of the analysis cell, degree of conversion of NO2 to NO and degradation of NO to nitrogen (N2) in the converter prior to analysis. Once correction factors were derived from calibration, the analyser was used to measure rate of decay of 40 ppm nitric oxide in both dry and humidified gases containing either 21%, 60% or 100% O2 in N2.ResultsCorrection factors obtained from calibration corrected falls in sensitivity to a maximum of 28%, variations in degree of conversion of NO2 to NO from 62-96% and degradation of NO from 2-8% in the converter before analysis depending on O2 concentration and humidification. When the calibrated machine was used to measure decay of NO, an excellent fit (r2 > 0.95) with a hyperbolic function was obtained confirming that rate of decay is proportional to oxygen O2 concentration and square of NO concentration. A rate constant per unit O2 concentration of 9.40E-10 ppm(-2) min(-1) for humidified gas was significantly higher than 8.27E-10 ppm(-2) min(-1) for "dry" gas (P = 0.008) at 22 degrees C. Rise in NO2 predicted from the "wet" rate constant achieved 3ppm in 65 seconds with 40 ppm NO in 100% oxygen and 107 sec. in 60% oxygen.ConclusionThis study indicates that for accurate analysis of NO and NO2 concentrations in airway gases a rigorous calibration of the instrument for the conditions applying in the experiment is mandatory. Once this is achieved the instrument is capable of precise analysis of both gases.

      Pubmed     Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…

Want more great medical articles?

Keep up to date with a free trial of metajournal, personalized for your practice.
1,694,794 articles already indexed!

We guarantee your privacy. Your email address will not be shared.