• Foot Ankle Int · Dec 2013

    Comparative Study

    Biomechanical comparison of syndesmotic injury fixation methods using a cadaveric model.

    • Edward Ebramzadeh, Ashleen R Knutsen, Sophia N Sangiorgio, Maximino Brambila, and Thomas G Harris.
    • J. Vernon Luck Research Center at Orthopaedic Hospital, Los Angeles, CA, USA.
    • Foot Ankle Int. 2013 Dec 1; 34 (12): 1710-7.

    BackgroundThere is growing interest in suture-button devices for syndesmosis injury, which are intended to offer less rigid fixation than screw fixation.MethodsThe fixation strength with 2 different suture-button devices, ZipTight and TightRope, were compared using 5 cadaveric leg pairs (n = 10). In an additional 5 pairs (n = 10), ZipTight was compared to 3.5 mm quadricortical screw fixation. Ankle motion was measured intact, then following simulated syndesmosis injury and fixation. Cyclic loads (peak 750 N, 7.5 Nm) were applied. Finally, external rotation to failure was measured and failure mode was documented.ResultsRange of motion increased after simulated injury and fixation with all devices (max 14.5 degrees). In all groups, diastasis remained below 1.0 mm intact and below 2.0 mm during cyclic loading. Compared to intact, under load to failure, diastasis with ZipTight devices increased by 4.7 ± 1.3 mm and 7.6 ± 4.3 mm, with TightRope, 6.3 mm, and screw construct, 1.3 mm. ZipTight specimens rotated approximately 80 ± 22 degrees before failure, TightRope, 67 ± 13 degrees, screw constructs, 76 ± 27 degrees. Mean failure torque was between 22.2 ± 6.9 Nm and 28.1 ± 12.7 Nm for ZipTight, compared to 32.9 ± 8.0 Nm for TightRope (P = .07), and 30.1 ± 9.6 Nm for screw constructs (P = .03). The majority of suture-button constructs failed by fibular fracture (ZipTight = 6, TightRope = 4), the remaining by device pull-through (ZipTight = 3, TightRope = 1) and loosening (ZipTight = 1). Conversely, 3 of screw-fixed specimens failed by device failure, 2 from bone fracture.ConclusionSuture-button devices provided torsional strength below that of screw fixation. However, all devices may provide failure torques well above 20 Nm, exceeding likely torques applied in casts during healing.(1,2,4) CLINICAL RELEVANCE: Suture-button devices appear to have provided adequate fixation strength for syndesmosis injuries.

      Pubmed     Full text   Copy Citation     Plaintext  

      Add institutional full text...

    Notes

     
    Knowledge, pearl, summary or comment to share?
    300 characters remaining
    help        
    You can also include formatting, links, images and footnotes in your notes
    • Simple formatting can be added to notes, such as *italics*, _underline_ or **bold**.
    • Superscript can be denoted by <sup>text</sup> and subscript <sub>text</sub>.
    • Numbered or bulleted lists can be created using either numbered lines 1. 2. 3., hyphens - or asterisks *.
    • Links can be included with: [my link to pubmed](http://pubmed.com)
    • Images can be included with: ![alt text](https://bestmedicaljournal.com/study_graph.jpg "Image Title Text")
    • For footnotes use [^1](This is a footnote.) inline.
    • Or use an inline reference [^1] to refer to a longer footnote elseweher in the document [^1]: This is a long footnote..

    hide…