Medical & biological engineering & computing
-
Med Biol Eng Comput · Sep 1995
Effects of electrode geometry and combination on nerve fibre selectivity in spinal cord stimulation.
The differential effects of the geometry of a rostrocaudal array of electrode contacts on dorsal column fibre and dorsal root fibre activation in spinal cord stimulation are analysed theoretically. 3-D models of the mid-cervical and mid-thoracic vertebral areas are used for the computation of stimulation induced field potentials, whereas a cable model of myelinated nerve fibre is used for the calculation of the excitation thresholds of large dorsal column and dorsal root fibres. The size and spacing of 2-D rectangular electrode contacts are varied while mono-, bi- and tripolar stimulation are applied. ⋯ Fibre type preference is most sensitive to variations of rostrocaudal contact size and least sensitive to variations of lateral contact size. Dorsal root fibre preference is increased and sensitivity to lead geometry is reduced as the distance from contacts to spinal cord is increased.