Clinical TrialA system for investigating oesophageal photoplethysmographic signals in anaesthetised patients.
The monitoring of arterial blood oxygen saturation in patients with compromised peripheral perfusion is often difficult, because conventional non-invasive techniques such as pulse oximetry (SpO2) can fail. Poor peripheral circulation commonly occurs after major surgery including cardiopulmonary bypass. The difficulties in these clinical situations might be overcome if the sensor were to monitor a better perfused central part of the body such as the oesophagus. ⋯ The ratio of the oesophageal to finger AC PPG amplitudes was calculated for the infrared and red wavelengths for each patient. The mean (+/- standard deviation) of this ratio was 2.9 +/- 2.1 (n = 19) for the infrared wavelength and 3.1 +/- 2.4 (n = 16) for the red wavelength. The red and infrared wavelengths used are appropriate for pulse oximetry and this investigation indicates that the mid-oesophagus may be a suitable site for the reliable monitoring of SpO2 in patients with poor peripheral perfusion.