Medical & biological engineering & computing
-
Med Biol Eng Comput · Jan 2005
Robust design of finger probe in non-invasive total haemoglobin monitor.
The development of a non-invasive monitor of total haemoglobin has been reported previously. The paper presents the design and testing of a finger probe used in that optical haemoglobin monitor. When a finger is inserted into the probe, light is radiated onto the nail, and a silicon detector measures transmitted light. ⋯ An optimally designed finger probe from DOE analysis, compared with the initial design, increased the correlation coefficient from 0.696 to 0.869 and improved the standard deviation from 1.18 to 0.81 g dl(-1) in predicting total haemoglobin. This was under different conditions of finger-probe alignment. Under different ambient light conditions, the optimum design improved the correlation coefficient from 0.735 to 0.870 and reduced the standard deviation from 1.14 to 0.83 g dl(-1).