Artificial organs
-
The objective of this study was to evaluate a pulsatile extracorporeal membrane oxygenation (ECMO) system in terms of hemodynamic energy generation and transmission under various pulsatile amplitudes, flow rates, and pseudopatient pressures in a simulated adult ECMO circuit. Surplus hemodynamic energy (SHE), a measure of the quality of pulsatility, was used to quantify pulsatile flow. The circuit consisted of an i-cor diagonal pump, an adult XLung oxygenator, a 21 Fr Medtronic Biomedicus femoral arterial cannula, a 23/25 Fr Sorin RAP femoral venous cannula, and 3/8 in ID tubing for both arterial and venous lines. ⋯ The novel pulsatile ECMO system can create nonpulsatile and pulsatile flow in an adult ECMO model. However, pulsatility gradually weakened with increasing flow rates. Pulsatile amplitude settings were found to have a great impact on pulsatility.