Pacing and clinical electrophysiology : PACE
-
Pacing Clin Electrophysiol · Dec 2003
Implantable cardioverter defibrillator events in patients with asymptomatic nonsustained ventricular tachycardia: is device implantation justified?
Primary prevention trials have demonstrated that patients with coronary disease, reduced left ventricular function, and nonsustained ventricular tachycardia (NSVT) have improved survival with implantable cardioverter defibrillator (ICD) therapy, presumably secondary to effective termination of life-threatening arrhythmias. However, stored intracardiac electrograms were not always available and specific arrhythmias leading to ICD therapy were not always known. We examined the occurrence of ICD events in 51 consecutive patients who match the described patient profile to determine the frequency of appropriate and inappropriate ICD therapy. ⋯ The mean arrhythmia rate leading to appropriate therapy for VT/VF was 232 +/- 72 beats/min (range 181-400 beats/min), and the mean rate leading to inappropriate therapy for SVT was 168 +/- 10 beats/min (range 160-180 beats/min). Patients with coronary disease and asymptomatic NSVT commonly receive appropriate defibrillator therapy. These results support the need for ICD implantation for primary prevention, with attention to careful programming of the detection rate to prevent inappropriate therapy.