International ophthalmology
-
To assess the intra-observer repeatability and inter-observer reproducibility of central corneal thickness (CCT) measurements of PachPen (Accutome, Inc., Pennsylvania, USA), a hand-held, portable ultrasonic pachymeter when used by an ophthalmic nurse compared to an ophthalmologist. Ophthalmology Clinic, University of Malaya Medical Center In this prospective study, CCT was measured in 184 eyes of 92 healthy subjects, first by a corneal surgeon experienced in ultrasound pachymetry (Observer 1) followed by an ophthalmic nurse new to the procedure (Observer 2). Nine measurements were obtained from each eye by each observer, independently. ⋯ The measurements by both observers showed high correlation (0.96) and good agreement (mean difference -2.4 µm; 95% limits of agreement -21.4, 16.7 µm). Coefficient of reproducibility of measurements between observers was 5.08%. Accutome PachPen hand-held ultrasound pachymeters gives excellent intra-observer repeatability and inter-observer reproducibility by personnel of different training grades.