Preclinical testing of femoral hip components: an experimental investigation with four prostheses.
Existing standards for the preclinical testing of femoral hip implants have been successful in the objective of guaranteeing the implant's fatigue strength. There is a need for an experimental test which could ensure prostheses were not susceptible to aseptic loosening. In this study we measure the relative movement between the prosthesis and the bone of four different cemented femoral component designs in in vitro tests. ⋯ With regard to the steady state inducible displacements of the prostheses, those of the Charnley, Exeter, and Lubinus decreased or were stable with respect to time, whilst those of the Müller typically increased with respect to time. It is concluded that migration is not a suitable basis for in vitro comparison of prosthesis designs. However inducible displacement trends provide a clinically comparable performance ranking.