Assessment of endotracheal tube placement in newborn infants: a randomized controlled trial.
International resuscitation guidelines recommend clinical assessment and exhaled CO2 to confirm tube placement immediately after intubation. However, exhaled CO2 devices can display false negative results. In comparison, any respiratory function monitor can be used to measure and display gas flow in and out of an endotracheal tube. However, neither method has been examined in detail. We hypothesized that a flow sensor would improve the assessment of tracheal vs esophageal tube placement in neonates with a higher success rate and a shorter time to tube placement confirmation when compared with the use of a quantitative end-tidal CO2 (ETCO2) detector. ⋯ A flow sensor would improve the assessment of successful endotracheal tube placement with a higher success rate and a shorter time compared with an ETCO2 detector.
-
To evaluate the accuracy of blood pressure (BP) measured non-invasively (NIBP) compared with invasive arterial BP (IABP) measured by umbilical arterial catheter in neonates undergoing therapeutic hypothermia (TH) for hypoxic ischemic encephalopathy (HIE). ⋯ In term or near-term neonates with HIE, NIBP measurements are a robust method to determine SBP; however, DBP and MBP are underestimated. Determination of hypotension using the MBP measured non-invasively should be interpreted with caution.