Computer methods and programs in biomedicine
-
Comput Methods Programs Biomed · May 2014
Impact of sensor and measurement timing errors on model-based insulin sensitivity.
A model-based insulin sensitivity parameter (SI) is often used in glucose-insulin system models to define the glycaemic response to insulin. As a parameter identified from clinical data, insulin sensitivity can be affected by blood glucose (BG) sensor error and measurement timing error, which can subsequently impact analyses or glycaemic variability during control. This study assessed the impact of both measurement timing and BG sensor errors on identified values of SI and its hour-to-hour variability within a common type of glucose-insulin system model. ⋯ Identified values of SI were mostly (90th percentile) within 29% of the true value when influenced by both sources of error. However, these effects may be overshadowed by physiological factors arising from the critical condition of the patients or other under-modelled or un-modelled dynamics. Thus, glycaemic control protocols that are designed to work with data from glucometers need to be robust to these errors and not be too aggressive in dosing insulin.